Lit M00502
Prism R™ Refrigerated Microcentrifuge 7 short-run of set speed. The short-run will stop if the button is released. Note: 1. For safet
8 4. Overview This section presents an overview of the Prism R microcentrifuge’s various components and control panel (see Figure 2) as wel
Prism R™ Refrigerated Microcentrifuge 9 Keypad Function Button Symbol Function Description START/STOP and Value Adjustment Knob Start/Stop a
10 5. Operation Users can pre-set the time, temperature and speed (RPM/RCF) before a run. Attention: Never attempt to operate the centri
Prism R™ Refrigerated Microcentrifuge 11 the display will flash. Turn the (START/STOP) knob to increase or decrease the value. Press button again
12 temperature can also be changed during the run. Pre-Cooling Pre-Cooling starts a temperature-control run at 6,000RPM. It allows the rotor
Prism R™ Refrigerated Microcentrifuge 13 5.6. Quick-Spin Operation The centrifuge can be operated for a short run by pressing and holdi
14 personnel only. Repairs performed by unauthorized personnel may void the warranty. 6.2. Cleaning the Centrifuge Always keep the cent
Prism R™ Refrigerated Microcentrifuge 15 To replace rotor, first make sure the motor shaft and rotor mounting hole are clean. Place the rotor on the
16 authorized service. Symptom Possible Cause Solution Press (START/STOP) knob when Lid is open. Close the lid “ ” Error Defective lid lock se
Prism R™ Refrigerated Microcentrifuge 17 Defective lid lock Open manually and have unit serviced No power from PC board Call for service Lid lo
18 Power Supply 100V/50~60 Hz, 6.0A 110V/ 50~60 Hz, 5.5A 230V/ 50~60 Hz, 2.8A Dimensions (W x L x H) 280 x 430 x 248 mm Weight (with Rotor) 22 Kgw
11 Declaration of Conformity Number: CE 00502 Manufacturer: Labnet International, Inc,., 31 Mayfield Ave., Edison, NJ 08837 USA Labnet
LIMITED WARRANTY Labnet International, Inc. warrants that this product will be free from defects in material and workmanship for a period of two (
Table of Content 1. Safety Precautions... 1 1.1. Operation Safety Precautions...
Prism R Refrigerated Microcentrifuge 1 1. Safety Precautions Before using the Prism R refrigerated microcentrifuge for the first time, pl
2 14. To guarantee sufficient ventilation, ensure that the centrifuge has at least 30 cm of free space on all sides, including the rear.
Prism R™ Refrigerated Microcentrifuge 3 2. General Description The Prism R refrigerated microcentrifuges is a compact, bench-top centrifuge w
4 Open the Prism R refrigerated microcentrifuge package and confirm that all items are included: Prism R unit with a 24 x 1.5/2ml rotor an
Prism R™ Refrigerated Microcentrifuge 5 3.3. Opening and Closing the Lid The lid can be opened only when the centrifuge is powered on. Press button
6 Figure 1. Loading the rotor 3.5. Starting and Stopping a Run Please refer Chapter 5 Operation to set up your preferred settings of time, speed